The versatile differential pressure Micromanometer ManoAir500 displays Pa, mbar and mmH2O for pressure as well as the flow velocity of gaseous media.

Thereby media temperature, effective ambient pressure and humidity are considered.
 
2 models are avaible:
Measuring range +/- 200 Pa or +/- 2000 Pa with automatic 10% measuring range changeover
Instrument features:
• 2 models: Measuring range +/- 200 Pa or +/- 2000 Pa with automatic 10% measuring range changeover
• Measurement units with changeover facility
Pressure Pa, mbar, mmH2O
Flow m/s, l/min, m3/h
Barometer kPa, bar
with humidity probe % rF
with temp. probe ºC
• Mean, minimum and maximum values per keystroke
• Automatic temperature compensation of pressure sensor
• Connection facility for external humidity and temperature probe.
• Thermocouples type E, J, K and T can be used in addition to NTC- and PT1000 - temperature probes
• Temperature, humidity and barometer readings available per keystroke
• Analogue output , Mini2Logger and RS232 connections
• Integrated barometer module
• Duct cross-section input facility per keystroke (for l/min or m3/h)
• Flow calculation based on standard or variable density
Technical specifications:
Effectve range Resolution Accuracy
ManoAir500 Mod. 1
100% range ± 200 Pa 0,1 Pa ± 0,2 Pa
10% range ± 20 Pa 0,01 Pa ± 0,05 Pa
ManoAir500 Mod. 2
100% range ± 2000 Pa 1 Pa ± 2 Pa
10% range ± 200 Pa 0,1 Pa ± 0,5 Pa
Barometer
(static pressure) 0 to 400 kPa (4 Bar) 1 kPa ± 1 kPa
Media dry instrument air or inert gases
Measurement units
Pressure: Pa, mbar, mmH2O
Flow: m/s, l/min, m3/h
Connector pins: push-on nipple for 3 mm int. dia. hose
Display LCD 4 digits
Supply Batteries (2 x 9 Volt) or external power supply
Output
Pressure: ± 1 Volt
Flow: ± 1 Volt
signal output for Mini2Logger (interval 0.5 sec)
signal output for RS232 (depending of chosen resol. 2 / 20sec)
Case dim. 190 x 150 x 70 mm (with carrying bracket 250 x 180 x 70 mm)
Case protection IP 40
Max.load 4 Bar (max. static pressure)
Humidity probe
Humidity: 0 - 99% rh 0,1% rh ± 3% rh
Temperature: - 20 to 60ºC 0,1 ºC ± 0,5 ºC
Universal 
temperature probe
- 20 to 140ºC 0,1 ºC
at 0 - 70ºC 0,2ºC
higher 70ºC 0,5ºC

Air 
temperature probe
- 20 to 140ºC 0,1 ºC
at 0 - 70ºC 0,2ºC
higher 70ºC 0,5ºC
Surface
temperature probe
- 20 to 140ºC 0,1 ºC
at 0 - 70ºC 0,2ºC
higher 70ºC 0,5ºC
Universal high
temperature probe
- 20 to 600ºC 0,1 ºC
at 0 - 70ºC 0,5ºC
higher 70ºC 1,0ºC
Thermocouple
with adapter
dependent on
thermocouple in use
1 ºC
dependent on
thermocouple in use
Specifications subject to change without further notice.